A meta-analysis published in Radiology reviews how artificial intelligence (AI) systems perform when they interpret digital mammography and digital breast tomosynthesis on their own. The results show diagnostic performance above that of radiologists in reader studies and in tomosynthesis, but not in historic cohorts, which calls for caution before treating AI as a full replacement for human reading.
What was already known
Interest in applying AI to mammographic screening keeps growing, but before it can be considered a modality able to interpret images independently its performance must be critically evaluated. Studies published so far varied widely in design, population and technology, making an overall conclusion difficult.
What this work adds
The authors ran a systematic search across PubMed, Google Scholar, Embase (Ovid) and Web of Science for studies published between January 2017 and June 2022. Sensitivity, specificity and area under the ROC curve (AUC) were reviewed, and methodological quality was assessed with the QUADAS-2 and QUADAS-C tools. Random-effects meta-analysis and meta-regression were performed overall and by study type (reader studies versus historic cohort studies) and by imaging technique (digital mammography versus tomosynthesis).
Main results
Sixteen studies were analysed, covering 1,108,328 examinations in 497,091 women: six reader studies, seven historic cohort studies on digital mammography and four studies on tomosynthesis. In the six digital mammography reader studies, pooled AUC was significantly higher for standalone AI than for radiologists (0.87 versus 0.81, P = .002), but no such difference appeared in historic cohort studies (0.89 versus 0.96, P = .152). In the four tomosynthesis studies, AUC was also significantly higher for AI than for radiologists (0.90 versus 0.79).
What it means
The meta-analysis suggests that standalone AI can reach diagnostic performance comparable to, or even better than, that of radiologists in certain settings, particularly in tomosynthesis and in controlled reader studies. The fact that this advantage does not carry over to historic cohorts is a reminder that results depend on study design and on the real-world setting where the tool is deployed. The available evidence still does not support replacing double reading.
A view from the breast unit
For breast units and screening programmes, these data point to AI as a potential support that could ease workload and improve detection, always within local validation and a quality-control framework, rather than as unsupervised replacement of radiological reading.
Reference: Standalone AI for Breast Cancer Detection at Screening Digital Mammography and Digital Breast Tomosynthesis: A Systematic Review and Meta-Analysis. Radiology, 2023. DOI